The Foxcroft FX-1000p-B Chlorine Analyzer utilizes the most accurate test method to determine chlorine residual levels in process water, a test similar to that used in lab environment titrator test methods. This makes it an excellent choice for your chlorine control or monitoring applications where the pH is steady. Residual readings are instantaneous and do not relay on "sample and hold" methods. this minimizes any delay in readings for control applications. | |
![]() |
![]() |
![]() |
![]() . |
Constantly measures free chlorine residuals in drinking water without a buffer solution, perfect for applications that have a steady pH sample. The FX-1000p-B Bufferless Chlorine Analyzer utilizes the most accurate test method to determine chlorine residuals levels in process water, similar to that used in lab environment titrator test methods. | |
![]() |
![]() |
![]() |
![]() |
![]() |
Monitor/ Control Well Pump Stations . |
![]() |
![]() |
Booster Pump Stations If you require to increase chlorine residuals, this application is prefect for you. By installing a sample point down stream from the chlorine ejection point you can both monitor and control. |
![]() |
|
![]() |
![]() |
![]() |
![]() |
Chlorine Booster Stations If you need to increase chlorine residual in remote areas, this application is perfect for you. By installing a sample pump down stream from the chlorine ejection point you can both monitor and control. By sizing the sample pump properly you can create a artificial flow condition, perfect in areas with no or very little flow conditions. Single point control is excellent for this application, a flow meter is not required. |
|
![]() |
![]() |
![]() |
Filter Plant Monitoring and Control Filter plant monitoring and control for chlorine residuals is very important, it ensures proper disinfections is taking place. If you are using the chlorine analyzer for control, be sure the chlorine injection point and the sample point are not to close to cause erratic readings and proper mixing of chlorine is taken place. |
![]() |
|
![]() |
![]() |
![]() |
A good process control system includes the following: • Properly placed sample point • Analyical Analyzer (chlorine, pH, etc.) • Single Loop Controller • Chemcial ejection system • Properly placed chemical injection point |
![]() |
![]() |
||
Process Time Interval Typically a good control interval would be somewhere between 2 and 10 minutes. If the control time interval is too long, any system upsets in residual will take a long time for the controller to correct. |
![]() ![]() ![]() FX-1000p-B Bufferless Chlorine Analyzer
Applications
|
|
Foxcroft Equipment and Service Company, Incorporated 2101 Creek Road - Glenmoore, PA 19343 - USA 1.800.874.0590 - 1.610.942.2888 - FAX 1.610.942.2769 Email: sales@foxcroft.com All Rights Reserved - All material contained in this Web Site is copyrighted - No material or ideas contained within this Web Site shall be used or copied without written consent from Foxcroft. ©2009 RWI3rd |