Location:

Mount the FX-1000p Chlorine Analyzer in a location that is:
Protected from elements (rain, snow, ice, dust, etc.)
Clean and dry where there is little or no vibration.
Protected from corrosive fluids.
Within ambient temperature limits (32-120°F, 0-49°C).

Important
Mounting the system is direct sunlight may increase the ambient temperature above the maximum limit.

Mounting:
Attach the four mounting ears to the rear of the remote enclosure. Anchor securely to a wall with the sample flow cell drain outlets pointing down. Mount the FX-1000p level to ensure that its integral weir operates properly, and at approximate shoulder height for convenience.